Radioactivity Retention of Liquid Iodine-131 in Residual Syringes: Implications for Radiation Exposure and Radioactive Waste Management in Radiopharmaceutical Preparation

Authors

  • Anggoro Praja Mukti Universitas Gadjah Mada
  • Hilda Ismail Universitas Gadjah Mada
  • Teuku Nanda Saifullah Sulaiman Universitas Gadjah Mada

DOI:

https://doi.org/10.59141/-.v8i2.561

Keywords:

iodine-131, radiopharmaceutical preparation, residual syringe activity, radioactivity retention, radiation exposure

Abstract

Residual radioactivity in syringes used during the preparation of liquid iodine-131 radiopharmaceuticals may contribute to avoidable radiation exposure and increase the radioactive waste burden. Although radioactivity retention in syringes has been reported for several radiopharmaceutical preparations, safety-oriented evaluations of residual iodine-131 activity in liquid formulations remain limited. This study aimed to investigate the relationship between residual radioactivity retention and radiation exposure rates in syringes following liquid iodine-131 preparation and to assess the implications for radioactive waste management. This experimental study used sodium iodide labeled with iodine-131 (Na¹³¹I) in liquid form. Initial and residual syringe activities were measured using a radionuclide dose calibrator, while ambient dose equivalent rates were measured using a gamma radiation survey meter. Residual activity, percentage radioactivity retention, radiation dose rate, and estimated decay-in-storage duration were analyzed descriptively. Pearson's correlation analysis and simple linear regression were performed to evaluate the relationship between radioactivity retention and measured radiation dose rates. Sequential rinsing with water for injection and 0.9% sodium chloride solution was evaluated as an approach to reducing residual radioactivity. Mean residual syringe activity increased from 0.270 ± 0.012 mCi at an initial activity concentration of 5 mCi/mL to 0.435 ± 0.016 mCi at 15 mCi/mL. Radioactivity retention increased from 12.90 ± 0.63% to 20.97 ± 0.74%, accompanied by an increase in the measured radiation dose rate from 2.703 ± 0.102 to 4.131 ± 0.237 µSv/h. Radioactivity retention exhibited a strong positive correlation with the radiation dose rate (Pearson's r = 0.9412, R² = 0.8858, p < 0.0001). Sequential rinsing reduced the measured dose rate from 4.160 ± 0.186 to 1.329 ± 0.034 µSv/h using water for injection and from 4.190 ± 0.117 to 1.314 ± 0.023 µSv/h using 0.9% sodium chloride solution. These findings demonstrate a strong association between residual syringe radioactivity and measured radiation dose rates following liquid iodine-131 preparation, highlighting the contribution of residual activity to the initial radioactive waste burden. Controlled rinsing procedures may help reduce residual syringe activity and support radiation protection, contamination control, and radioactive waste management practices in nuclear medicine facilities.

Downloads

Download data is not yet available.

References

Barbosa, N., Sandoval Castillo, L., & Quimbayo, J. S. (2022). Discharges of Nuclear Medicine Radioisotopes: The Impact of an Abatement System. Health Physics, 122(5), 586–593. https://doi.org/10.1097/HP.0000000000001543

Bergeron, D. E., & Cessna, J. T. (2018). An update on “dose calibrator” settings for nuclides used in nuclear medicine. Nuclear Medicine Communications, 39(6), 500–504. https://doi.org/10.1097/MNM.0000000000000833

Bly, R. (2023). Radiation safety of current European practices of therapeutic nuclear medicine: survey results from 20 HERCA countries. Journal of Radiological Protection, 43(1). https://doi.org/10.1088/1361-6498/acafef

Galbraith, W., Chen, X., Talley, K., & Grantham, V. (2015). Assessment of 99mTc-succimer residual activity using inert nonreactive syringes. Journal of Nuclear Medicine Technology, 43(1), 61–63. https://doi.org/10.2967/jnmt.114.147983

Gillings, N., Hjelstuen, O., Ballinger, J., Behe, M., Decristoforo, C., Elsinga, P., Ferrari, V., Peitl, P. K., Koziorowski, J., Laverman, P., Mindt, T. L., Neels, O., Ocak, M., Patt, M., & Todde, S. (2021). Guideline on current good radiopharmacy practice (cGRPP) for the small-scale preparation of radiopharmaceuticals. EJNMMI Radiopharmacy and Chemistry, 6(1), 8. https://doi.org/10.1186/s41181-021-00123-2

Gunasekera, R. D., Notghi, A., Mostafa, A. B., & Harding, L. K. (2001). Adsorption of radiopharmaceuticals to syringes leads to lower administered activity than intended. Nuclear Medicine Communications, 22(5), 493–497. https://doi.org/10.1097/00006231-200105000-00005

Haugen, B. R., Alexander, E. K., Bible, K. C., Doherty, G. M., Mandel, S. J., Nikiforov, Y. E., Pacini, F., Randolph, G. W., Sawka, A. M., Schlumberger, M., Schuff, K. G., Sherman, S. I., Sosa, J. A., Steward, D. L., Tuttle, R. M., & Wartofsky, L. (2016). 2015 American Thyroid Association Management Guidelines for Adult Patients with Thyroid Nodules and Differentiated Thyroid Cancer. Thyroid, 26(1), 1–133. https://doi.org/10.1089/thy.2015.0020

Hung, J. C. (1997). Radioiodine dispensing and usage in a centralized hospital nuclear pharmacy. Thyroid, 7(2), 289–294. https://doi.org/10.1089/thy.1997.7.289

Keskintepe, D., & Özer, A. Y. (2005). Residual radioactivity in the syringes used in injection of radiopharmaceuticals and factors affecting the amount of residual radioactivity. Fabad Journal of Pharmaceutical Sciences, 30(4), 176–180.

Kvaternik, H., Gatterer, J., Plhak, E., Schwarzgruber, J. F., & Aigner, R. M. (2020). Systematic Assessment of the Adsorption of 99mTc-Radiopharmaceuticals onto Plastic Syringes. Journal of Nuclear Medicine Technology, 48(2), 168–173. https://doi.org/10.2967/jnmt.119.235432

Ladrière, T., Aussage, L., & Vigne, J. (2024). Container-content interactions with radiopharmaceuticals: Seeing is believing. European Journal of Pharmaceutics and Biopharmaceutics, 196, 114200. https://doi.org/10.1016/j.ejpb.2024.114200

Ladriere, T., Desmonts, C., Zimmermann, P.-A., Peyronnet, D., Agostini, D., & Vigne, J. (2023). Impact of disposable syringes type choice on myocardial perfusion imaging procedures with [99mTc]Tc-tetrofosmin. Journal of Nuclear Cardiology, 30(4), 1399–1405. https://doi.org/10.1007/s12350-022-03143-w

Mushtaq, A., Ur Rehman, T., Safdar Mansur, M., & Jehangir, M. (2008). Adsorption of 99mTc-radiopharmaceuticals onto injection vials and syringes. Journal of Nuclear Medicine Technology, 36(2), 91–94. https://doi.org/10.2967/jnmt.107.048561

Nguyen, N. C., Anigati, E. M., Desai, N. B., & Öz, O. K. (2024). Radioactive iodine therapy in differentiated thyroid cancer: An update on dose recommendations and risk of secondary primary malignancies. Seminars in Nuclear Medicine, 54(4), 488–496. https://doi.org/10.1053/j.semnuclmed.2024.05.002

Saha, G. B. (2004). Fundamentals of Nuclear Pharmacy (5th ed.). Springer.

Stavrou, P. Z., Papachristou, M., Persakis, E., Kouvelis, K., & Datseris, I. E. (2016). Residual activities of 99mTc-labelled radiopharmaceuticals in routine nuclear medicine practice. Nuclear Medicine Communications, 37(6), 658–663. https://doi.org/10.1097/MNM.0000000000000487

Stopar, T. G., Socan, A., & Peitl, P. K. (2007). Adsorption of radiopharmaceuticals to syringes: Setting up a reliable protocol for its assessment. Nuclear Medicine Communications, 28(12), 951–955. https://doi.org/10.1097/MNM.0b013e3282f1b9b7

Swanson, T. N., Troung, D. T., Paulsen, A., Hruska, C. B., & O’Connor, M. K. (2013). Adsorption of 99mTc-sestamibi onto plastic syringes: Evaluation of factors affecting the degree of adsorption and their impact on clinical studies. Journal of Nuclear Medicine Technology, 41(4), 247–252. https://doi.org/10.2967/jnmt.113.132159

Wadsley, J., Armstrong, N., Bassett-Smith, V., Beasley, M., Chandler, R., Cluny, L., Craig, A. J., Farnell, K., Garcez, K., Garnham, N., Graham, K., Hallam, A., Hill, S., Hobrough, H., McKiddie, F., & Strachan, M. W. J. (2023). Patient preparation and radiation protection guidance for adult patients undergoing radioiodine treatment for thyroid cancer in the UK. Clinical Oncology, 35(1), 42–56. https://doi.org/10.1016/j.clon.2022.07.002

Willegaignon, J., Fernandes, S. C. P., Pelissoni, R. A., Coura-Filho, G. B., Sapienza, M. T., & Buchpiguel, C. A. (2023). Radiation safety measures in diagnostic nuclear medicine, based on the potential radiation dose emitted by radioactive patients. Radiologia Brasileira, 56(1), 13–20. https://doi.org/10.1590/0100-3984.2022.0064

Downloads

Published

2026-10-15

How to Cite

Mukti, A. P., Ismail , H., & Saifullah Sulaiman , T. N. (2026). Radioactivity Retention of Liquid Iodine-131 in Residual Syringes: Implications for Radiation Exposure and Radioactive Waste Management in Radiopharmaceutical Preparation. Jurnal Sehat Indonesia (JUSINDO), 8(2), 906–915. https://doi.org/10.59141/-.v8i2.561